A high-risk medical device is assessed within the MDR procedure, and the MDR class under Rule 11 decides if it is high-risk.
Key findings
Article 43(3) puts the AI Act assessment of a device high-risk under Article 6(1) inside the MDR procedure: one technical file, one quality management system, one declaration of conformity.
The MDR Rule 11 class decides: AI software at Class IIa or above is high-risk, while Class I software is outside Article 6(1), though not outside the Act.
Existing medical devices notified bodies may assess the AI Act requirements under a transitional power, but must apply for designation under the AI Act by 28 January 2028.
In the paper
One procedure, by Article 43(3)
Rule 11 decides whether the device is high-risk
One procedure does not guarantee one notified body
The relief is procedural; the requirements are not